We thinks what customers think, the urgency of urgency to act during the interests of a purchaser position of theory, allowing for much better good quality, lower processing costs, prices are extra reasonable, won the new and old buyers the support and affirmation for 6 Cfm Vacuum Pump, Electric driven gate valve, Best Hvac Vacuum Pump, We have now designed a reputable track record among many shoppers. Quality&customer initially are normally our constant pursuit. We spare no attempts to produce greater solutions. Stay up for long-term cooperation and mutual positive aspects!
Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY Detail:

The ZQJ-3200 Helium Leak Detectors are microprocessor-controller leak detecting instruments. All the processes in the instrument are controlled automatically.

Advanteges:

1. Easy operation–Small volume, light weight, compact structure, remote operation panel

2. Multiful interface–Rs232, Digital I/O, USD port

3. Powerful functions–Various test mode, capability of detecting H2. 3He, amd 4He, multiple menu setting

4. Reliable performance–High sensitivity, wide measurement range, high inlet pressure, fast response time

5. Reliable quality–Prolong the service life, resistance of yttrium oxide iridium filament

Specifications:

Type ZQJ-3200
Smallest Detectable Leak Rate(Pa•m3/s) 5×10-13  vacuum mode5×10-10  sniffing mode
Leak Rate Display(Pa•m3/s) 10-13~10-1
Maximum Inlet Pressure(Pa) 2500
Response Time(s) ≤2
Run-up Time(min) <3
Power 230 VAC±10%/50 Hz
120V±10%/60 Hz,10A
Working Temperature and Relative humidity Working Temperature 10~35℃,realtive humidity ≤80%
L*W*H(mm) 550×460×304
Weight(kg) 44

Methods:

Vacuum Method

In the vacuum method test gas is blown against the wall of the evacuated sample from the atmosphere side. It enters the sample at leaks and is fed to the leak detector.

The sample must be vacuum pressure-proof.

The sensitivity stages GROSS —FINE—ULTRA are run through.

The detection limit is lower than in the sniffing method. The helium concentration at the leak must be known in order to quantify the leak. The state of equilibrium must be waited for.

Sniffing Method

In the sniffing method the test gas escaping from leaks in the sample into the atmosphere is detected.

The sample must withstand the applied test pressure.

In operation with the sniffing probe a constant gas flow is sucked in from the atmosphere. The helium proportion of the air (5.2 ppm) causes a leak rate display of approx. 1*10-6 mbar l/s which can be eliminated by the ZERO function.20 3 Description Operating Instructions, ikna88en1-01, 1605

To detect a leak, the sniffing probe is applied to the points of the sample under helium overpressure which are suspected of leaking. An increased leak rate value indicates an increased concentration of helium and therefore a leak. The higher the pressure and the helium concentration in the sample, the smaller the leaks which can be detected.

The sensitivity stages GROSS—FINE are run through.

The detection sensitivity and the quantifiability of the leak rate are less favourable than in the vacuum pressure leak detection.



Product detail pictures:

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures


Related Product Guide:

Along with the "Client-Oriented" small business philosophy, a rigorous high-quality handle system, highly developed producing machines and a powerful R&D group, we always supply high-quality products and solutions, fantastic services and aggressive costs for Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY , The product will supply to all over the world, such as: Latvia, Honduras, Angola, Our company will continue to adhere to the " superior quality, reputable, the user first " principle wholeheartedly. We warmly welcome friends from all walks of life to visit and give guidance, work together and create a brilliant future!
  • The supplier cooperation attitude is very good, encountered various problems, always willing to cooperate with us, to us as the real God.
    5 Stars By Betty from Sheffield - 2018.12.10 19:03
    The customer service staff's attitude is very sincere and the reply is timely and very detailed, this is very helpful for our deal,thank you.
    5 Stars By Fernando from Ottawa - 2018.11.28 16:25
    Write your message here and send it to us