owing to excellent assistance, a variety of top of the range items, aggressive costs and efficient delivery, we take pleasure in a very good standing amongst our shoppers. We've been an energetic corporation with wide market for Gate Valve Types, Glycerine Filled Pressure Gauge, Helium Leak Test Equipment, We invites you and your enterprise to thrive together with us and share a bright future in global market.
Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY Detail:

The ZQJ-3200 Helium Leak Detectors are microprocessor-controller leak detecting instruments. All the processes in the instrument are controlled automatically.

Advanteges:

1. Easy operation–Small volume, light weight, compact structure, remote operation panel

2. Multiful interface–Rs232, Digital I/O, USD port

3. Powerful functions–Various test mode, capability of detecting H2. 3He, amd 4He, multiple menu setting

4. Reliable performance–High sensitivity, wide measurement range, high inlet pressure, fast response time

5. Reliable quality–Prolong the service life, resistance of yttrium oxide iridium filament

Specifications:

Type ZQJ-3200
Smallest Detectable Leak Rate(Pa•m3/s) 5×10-13  vacuum mode5×10-10  sniffing mode
Leak Rate Display(Pa•m3/s) 10-13~10-1
Maximum Inlet Pressure(Pa) 2500
Response Time(s) ≤2
Run-up Time(min) <3
Power 230 VAC±10%/50 Hz
120V±10%/60 Hz,10A
Working Temperature and Relative humidity Working Temperature 10~35℃,realtive humidity ≤80%
L*W*H(mm) 550×460×304
Weight(kg) 44

Methods:

Vacuum Method

In the vacuum method test gas is blown against the wall of the evacuated sample from the atmosphere side. It enters the sample at leaks and is fed to the leak detector.

The sample must be vacuum pressure-proof.

The sensitivity stages GROSS —FINE—ULTRA are run through.

The detection limit is lower than in the sniffing method. The helium concentration at the leak must be known in order to quantify the leak. The state of equilibrium must be waited for.

Sniffing Method

In the sniffing method the test gas escaping from leaks in the sample into the atmosphere is detected.

The sample must withstand the applied test pressure.

In operation with the sniffing probe a constant gas flow is sucked in from the atmosphere. The helium proportion of the air (5.2 ppm) causes a leak rate display of approx. 1*10-6 mbar l/s which can be eliminated by the ZERO function.20 3 Description Operating Instructions, ikna88en1-01, 1605

To detect a leak, the sniffing probe is applied to the points of the sample under helium overpressure which are suspected of leaking. An increased leak rate value indicates an increased concentration of helium and therefore a leak. The higher the pressure and the helium concentration in the sample, the smaller the leaks which can be detected.

The sensitivity stages GROSS—FINE are run through.

The detection sensitivity and the quantifiability of the leak rate are less favourable than in the vacuum pressure leak detection.



Product detail pictures:

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures


Related Product Guide:

We have state-of-the-art equipment. Our products are exported for the USA, the UK and so on, enjoying a fantastic status among the clients for Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY , The product will supply to all over the world, such as: Turkey, Montreal, Ghana, Immediate and specialist after-sale service supplied by our consultant group has happy our buyers. Detailed Info and parameters from the merchandise will probably be sent to you for any thorough acknowledge. Free samples may be delivered and company check out to our corporation. n Morocco for negotiation is constantly welcome. Hope to get inquiries type you and construct a long-term co-operation partnership.
  • Good quality and fast delivery, it's very nice. Some products have a little bit problem, but the supplier replaced timely, overall, we are satisfied.
    5 Stars By Norma from Mecca - 2018.06.05 13:10
    The company has a good reputation in this industry, and finally it tured out that choose them is a good choice.
    5 Stars By Brook from Philippines - 2018.12.22 12:52
    Write your message here and send it to us