Good Quality Helium Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY

Short Description:


Product Detail

Product Tags

Related Video

Feedback (2)

"Sincerity, Innovation, Rigorousness, and Efficiency" could be the persistent conception of our organization for your long-term to establish alongside one another with shoppers for mutual reciprocity and mutual benefit for 1 Bar Vacuum Gauge, Vane Air Pump, Micro Pressure Gauge, Welcome you to be a part of us alongside one another to create your company easier. We're normally your finest partner when you want to have your own organization.
Good Quality Helium Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY Detail:

The ZQJ-3200 Helium Leak Detectors are microprocessor-controller leak detecting instruments. All the processes in the instrument are controlled automatically.

Advanteges:

1. Easy operation–Small volume, light weight, compact structure, remote operation panel

2. Multiful interface–Rs232, Digital I/O, USD port

3. Powerful functions–Various test mode, capability of detecting H2. 3He, amd 4He, multiple menu setting

4. Reliable performance–High sensitivity, wide measurement range, high inlet pressure, fast response time

5. Reliable quality–Prolong the service life, resistance of yttrium oxide iridium filament

Specifications:

Type ZQJ-3200
Smallest Detectable Leak Rate(Pa•m3/s) 5×10-13  vacuum mode5×10-10  sniffing mode
Leak Rate Display(Pa•m3/s) 10-13~10-1
Maximum Inlet Pressure(Pa) 2500
Response Time(s) ≤2
Run-up Time(min) <3
Power 230 VAC±10%/50 Hz
120V±10%/60 Hz,10A
Working Temperature and Relative humidity Working Temperature 10~35℃,realtive humidity ≤80%
L*W*H(mm) 550×460×304
Weight(kg) 44

Methods:

Vacuum Method

In the vacuum method test gas is blown against the wall of the evacuated sample from the atmosphere side. It enters the sample at leaks and is fed to the leak detector.

The sample must be vacuum pressure-proof.

The sensitivity stages GROSS —FINE—ULTRA are run through.

The detection limit is lower than in the sniffing method. The helium concentration at the leak must be known in order to quantify the leak. The state of equilibrium must be waited for.

Sniffing Method

In the sniffing method the test gas escaping from leaks in the sample into the atmosphere is detected.

The sample must withstand the applied test pressure.

In operation with the sniffing probe a constant gas flow is sucked in from the atmosphere. The helium proportion of the air (5.2 ppm) causes a leak rate display of approx. 1*10-6 mbar l/s which can be eliminated by the ZERO function.20 3 Description Operating Instructions, ikna88en1-01, 1605

To detect a leak, the sniffing probe is applied to the points of the sample under helium overpressure which are suspected of leaking. An increased leak rate value indicates an increased concentration of helium and therefore a leak. The higher the pressure and the helium concentration in the sample, the smaller the leaks which can be detected.

The sensitivity stages GROSS—FINE are run through.

The detection sensitivity and the quantifiability of the leak rate are less favourable than in the vacuum pressure leak detection.



Product detail pictures:

Good Quality Helium Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Good Quality Helium Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Good Quality Helium Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Good Quality Helium Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Good Quality Helium Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Good Quality Helium Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures


Related Product Guide:

It is actually our accountability to satisfy your needs and effectively serve you. Your pleasure is our best reward. We're on the lookout forward for your stop by for joint growth for Good Quality Helium Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY , The product will supply to all over the world, such as: Bolivia, California, South Africa, We warmly welcome your patronage and will serve our clients both at home and abroad with products and solutions of superior quality and excellent service geared to the trend of further development as always. We believe you will benefit from our professionalism soon.
  • This manufacturer can keep improving and perfecting products and service, it is in line with the rules of market competition, a competitive company.
    5 Stars By Carol from Netherlands - 2017.06.29 18:55
    In China, we have many partners, this company is the most satisfying to us, reliable quality and good credit, it is worth appreciation.
    5 Stars By Nicola from Gambia - 2017.04.18 16:45
    Write your message here and send it to us